Description
» Miris cleaner is offered as a cleaning solution for the Miris Human Milk Analyzer.
» Application: Only for cleaning of the Miris HMA cuvette and surface of the instrument. For in vitro diagnostic use.
» Storage: Store the concentrate at room temperature (20-30°C) in a sealed container. Store the diluted solution in glass or plastic containers, out of direct sunlight. Keep away from foodstuff and animal feed.
» Instructions: Dilute 50 ml concentrate (1 tube) with 950 ml distilled or deionized water and mix. Warm to 40° C before use and follow the instructions as set out.
» Clean instruments in- and outlets frequently. Minimum ounce a week or after every 100th sample, whichever come first.
» 5 tubes of 50ml
Miris is a company who specialize in human milk analysis of pre-term babies. They believe by knowing the macro nutrient composition of human milk, it can be determined whether milk fortification is needed.
Miris is operating on all continents and HMA™ is registered as a Medical Device in Europe, Japan and Indonesia.
Click here to watch a short video on how to mix the cleaner solution. To follow the cleaning procedure, click here.